Improved acoustic coupling for shock wave lithotripsy.

Practice Point

Joshua S Neucks, Yuri A Pishchalnikov, Anthony J Zancanaro, Jonathan N VonDerHaar, James C Williams, James A McAteer

Urological research · 2008 · Vol. 36 · Issue 1 · 61-6

PMID: 18172634 DOI: 10.1007/s00240-007-0128-y

Academy Summary

Neucks and colleagues investigated a very practical question in shock wave lithotripsy:

How should coupling gel actually be applied to minimise trapped air and improve shock-wave transmission?

Previous experimental work had already demonstrated that air pockets at the interface between the lithotripter treatment head and the patient can markedly reduce transmission of shock-wave energy and impair stone fragmentation.

This study therefore moved beyond identifying the problem and attempted to determine a practical coupling technique that operators could use during SWL.

The experiments were performed using a Dornier DoLi-50 electromagnetic lithotripter.

LithoClear coupling gel was used between the lithotripter treatment-head cushion and the acoustic window of a transparent test tank representing the patient.

Different methods of applying coupling gel were tested.

The amount of trapped air was then quantified using digital imaging, and different coupling techniques were tested for their effect on gypsum model-stone fragmentation.

The method of gel application had a major effect on coupling quality.

When gel was dispensed from a squeeze bottle and spread manually, coupling defects occupied approximately:

26.5 ± 2.7% of the interface

In contrast, when a large volume of approximately 250 mL of gel was poured directly from the stock container as a mound onto the treatment head and allowed to spread naturally when contact was established, coupling defects were reduced to:

1.2 ± 0.5%

This difference was highly significant.

Stone fragmentation was also significantly more efficient when coupling was established using the large-volume bolus technique rather than manually spreading the gel.

The investigators concluded that excessive handling of the coupling medium promotes the introduction of air bubbles.

They therefore recommended:

  • Applying a generous bolus of coupling gel
  • Applying it primarily to the treatment head
  • Avoiding unnecessary manual spreading
  • Allowing the gel to spread naturally as the therapy head contacts the patient

The study also demonstrated that inflation of the lithotripter water cushion could help collapse existing air pockets and improve poor coupling.

However, this was not as effective as establishing good coupling correctly from the beginning.

The fundamental principle is:

Good coupling begins with how the gel is applied.

Current EAU guidance continues to recognise acoustic coupling as an important determinant of SWL performance and gives a strong recommendation to ensure correct use of coupling agent because effective shock-wave transmission depends on it.

Unlock the Academy Interpretation for Free

Create a free Lithotripsy Academy account or log in to access the complete Academy interpretation, including Key Findings, Clinical Takeaway, Limitations, and Expert Commentary

Subscribe to View Details Already a member? Log in
Comments & Discussion0 comments

Discuss the publication's evidence, limitations and practical relevance. Please do not include identifiable patient information or individual medical advice.

No comments yet. Start the discussion.