A clinical observational study of effectiveness of a solid coupling medium in extracorporeal shock wave lithotripsy.

Practice Point

Hao-Han Chang, Yu-Chih Lin, Ching-Chia Li, Wen-Jeng Wu, Wen-Chin Liou, Yusen Eason Lin, Kuo-Kuang Huang, Wei-Chuan Chen

BMC urology · 2022 · Vol. 22 · Issue 1 · 56

PMID: 35413871 DOI: 10.1186/s12894-022-01001-y

Academy Summary

Chang and colleagues evaluated whether a solid coupling medium, termed the isolation coupling pad or icPad, could improve clinical stone disintegration during extracorporeal shock wave lithotripsy (ESWL) compared with conventional semi-liquid coupling gel.

This study is particularly relevant because it represents the clinical continuation of earlier phantom research showing that the icPad substantially reduced trapped air at the coupling interface.

The study included 300 patients with renal or ureteral stones treated between October 2017 and May 2018 using an electromagnetic lithotripter.

Patients were divided into three groups:

  • Group C: 103 patients treated using conventional semi-liquid gel
  • Group P1: 100 patients treated using icPad with 3,000 shock waves
  • Group P2: 97 patients treated using icPad with 2,500 shock waves

Shock-wave frequency was 90 shocks/min in all three groups. The published treatment table reports energy level 3 for all three groups.

The groups were well balanced for age, stone size, side and anatomical location.

Mean stone size was approximately:

  • 8.14 mm in Group C
  • 7.94 mm in Group P1
  • 8.14 mm in Group P2

Stones included:

  • Renal stones
  • Upper ureteral stones
  • Middle ureteral stones
  • Lower ureteral stones

The investigators defined successful stone disintegration as either:

  • Complete absence of stone fragments

or

  • Residual stone burden <4 mm

on follow-up imaging within 90 days.

Using this definition, successful stone disintegration rates were:

  • Semi-liquid gel Group C: 55.3%
  • icPad Group P1: 73.0%
  • icPad Group P2: 73.2%

The difference between Group P1 and conventional gel was statistically significant:

73.0% vs. 55.3%, p = 0.009

There was no significant difference between the two icPad groups:

73.0% vs. 73.2%, p = 0.975

despite Group P2 receiving only 2,500 shock waves compared with 3,000 in Group P1.

No major organ complications were reported in any group.

The study therefore provides clinical support for the concept that:

Improving the coupling interface may improve SWL effectiveness without simply increasing shock-wave number.

Current EAU guidance strongly supports the broader principle. Proper acoustic coupling between the treatment head and the patient's skin is considered important, and correct use of the coupling agent carries a strong recommendation because it is crucial for effective shock-wave transmission.

Unlock the Academy Interpretation for Free

Create a free Lithotripsy Academy account or log in to access the complete Academy interpretation, including Key Findings, Clinical Takeaway, Limitations, and Expert Commentary

Subscribe to View Details Already a member? Log in
Comments & Discussion0 comments

Discuss the publication's evidence, limitations and practical relevance. Please do not include identifiable patient information or individual medical advice.

No comments yet. Start the discussion.