Courses List

0
No products in the cart.
← Literature Database

Monitoring the coupling of the lithotripter therapy head with skin during routine shock wave lithotripsy with a surveillance camera.

Practice Point

Christian Bohris, Alexander Roosen, Martin Dickmann, Yasmin Hocaoglu, Stefan Sandner, Markus Bader, Christian G Stief, Sebastian Walther

The Journal of urology · 2012 · Vol. 187 · Issue 1 · 157-63

PMID: 22100005 DOI: 10.1016/j.juro.2011.09.039

Academy Summary

Bohris and colleagues investigated the quality of acoustic coupling during routine clinical shock wave lithotripsy (SWL) and examined how air within the coupling interface affects stone fragmentation efficiency.

The study addressed an important practical problem with modern dry-head lithotripters.

Shock waves must travel from the treatment head through:

  • Water-filled therapy cushion
  • Coupling gel
  • Patient's skin
  • Soft tissue
  • Stone

Air trapped between the treatment head and the patient's skin can significantly interfere with this acoustic pathway.

The investigators integrated a video surveillance camera into a Dornier DoLi SII lithotripter, allowing continuous visualization of the coupling interface during 30 routine SWL treatments.

Importantly, the treating urologist was blinded to the camera image, reproducing normal clinical practice in which the coupling interface is usually invisible to the operator.

The clinical component included:

  • 30 SWL treatment sessions
  • 26 patients
  • 21 men
  • 5 women
  • Mean BMI approximately 27.2 kg/m²
  • 26 renal stone treatments
  • 4 ureteral stone treatments

Four urologists performed the treatments.

Stone localisation was performed using:

  • X-ray in 20 treatments
  • Isocentric ultrasound in 10 treatments

Three different coupling gels were evaluated:

  • LithoClear
  • Sonogel
  • A custom-made low-viscosity gel

Good coupling, defined as less than 5% air within the relevant coupling area, was achieved in only:

10 of 30 treatments

In contrast:

  • 20 of 30 treatments had more than 5% air
  • 8 treatments had more than 20% air

The low-viscosity gel produced the best coupling conditions.

The investigators then tested the effect of different amounts of air on fragmentation using model stones.

The mean number of shock waves required for complete fragmentation was:

  • 0% air: 100 ± 4 shocks
  • 5% air: 126 ± 3 shocks
  • 10% air: 151 ± 8 shocks
  • 20% air: 287 ± 5 shocks

Therefore, when approximately 20% of the coupling area contained air, almost three times as many shock waves were required to achieve the same model-stone fragmentation.

The study demonstrated that poor coupling is not an occasional theoretical problem.

It occurs commonly during routine SWL and can substantially reduce the efficiency of energy transmission.

Current EAU guidance reflects this principle and states that proper acoustic coupling is important for SWL outcome, with a strong recommendation to ensure correct use of the coupling agent because it is crucial for effective shock-wave transmission.

Unlock the Academy Interpretation for Free

Create a free Lithotripsy Academy account or log in to access the complete Academy interpretation, including Key Findings, Clinical Takeaway, Limitations, and Expert Commentary

Subscribe to View Details Already a member? Log in
Comments & Discussion0 comments

Discuss the publication's evidence, limitations and practical relevance. Please do not include identifiable patient information or individual medical advice.

No comments yet. Start the discussion.