Courses List

0
No products in the cart.
← Literature Database

The effects of extracorporeal shock wave lithotripsy on pacemaker function.

Landmark Study

J Langberg, J Abber, J W Thuroff, J C Griffin

Pacing and clinical electrophysiology : PACE · 1987 · Vol. 10 · Issue 5 · 1142-6

PMID: 2444938 DOI: 10.1111/j.1540-8159.1987.tb06133.x

Academy Summary

Langberg and colleagues performed an early experimental investigation into the effects of extracorporeal shock wave lithotripsy (SWL) on cardiac pacemaker function.

The study specifically addressed two concerns that were important during the early introduction of SWL:

  • Could shock waves physically damage or reprogram a pacemaker?
  • Could shock-wave delivery interfere temporarily with pacemaker sensing or pacing output?

The investigators exposed 22 pacemaker pulse generators to shock waves generated by a lithotripter.

The pacemaker generator and distal portion of the pacing lead were positioned approximately 5 cm from the lithotripter focal point, with approximately 10 cm separating the generator and lead. Pacemaker behaviour was then assessed under different shock-wave delivery conditions. (pubmed.ncbi.nlm.nih.gov)

When shock waves were delivered synchronously with pacemaker output, only one of the 22 devices demonstrated abnormal behaviour, intermittently reverting to its magnet rate.

However, when shock waves were delivered at a rate faster than the pacemaker escape rate, approximately 50% of the devices demonstrated temporary inhibition caused by electromechanical interference.

Both:

  • Unipolar pacemakers
  • Bipolar pacemakers

were affected. (pubmed.ncbi.nlm.nih.gov)

Importantly, post-exposure analysis demonstrated that none of the pacemakers sustained permanent damage or inappropriate reprogramming.

The investigators therefore concluded that shock-wave exposure itself did not appear to permanently damage pacemaker generators, but inappropriate sensing and temporary pacing inhibition could occur when shock-wave delivery was not appropriately synchronized.

Their central practical conclusion was that the likelihood of false pacemaker inhibition appeared very low when shock waves were delivered synchronously with the QRS complex. (pubmed.ncbi.nlm.nih.gov)

This study provided some of the earliest mechanistic evidence supporting the transition from:

“Pacemaker = contraindication to SWL”

toward:

“Pacemaker = SWL may be possible with appropriate synchronization and device precautions.”

Unlock the Academy Interpretation for Free

Create a free Lithotripsy Academy account or log in to access the complete Academy interpretation, including Key Findings, Clinical Takeaway, Limitations, and Expert Commentary

Subscribe to View Details Already a member? Log in

Continue Learning

Explore related clinical guidance and structured education.

Comments & Discussion0 comments

Discuss the publication's evidence, limitations and practical relevance. Please do not include identifiable patient information or individual medical advice.

No comments yet. Start the discussion.